CVE-2025-54304
Description
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
CVSS Vector Breakdown
AV:NAttack VectorAC:LAttack ComplexityPR:NPrivileges RequiredUI:NUser InteractionS:UScopeC:HConfidentialityI:HIntegrityA:HAvailabilityWeaknesses
Affected Products
Exploitability
Attack Graph
Click technique nodes for MITRE ATT&CK details · drag to pan · Ctrl/⌘ + scroll to zoom, or go fullscreen.
MITRE ATT&CK
1 techniqueReferences
Unlock Complete Vulnerability Intelligence
Get the full picture for CVE-2025-54304 and every CVE in our database. Create a free account — no credit card required.
Create Free Account