CVE-2018-8868
Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function
Description
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
CVSS Vector Breakdown
AV:PAttack VectorAC:HAttack ComplexityPR:LPrivileges RequiredUI:NUser InteractionS:CScopeC:HConfidentialityI:LIntegrityA:LAvailabilityWeaknesses
Affected Products
Exploitability
References
Timeline
Unlock Complete Vulnerability Intelligence
Get the full picture for CVE-2018-8868 and every CVE in our database. Create a free account — no credit card required.
Create Free Account