CVE Tools

CVE-2018-10601

CVE-2018-10601 is a stack overflow in Philips Intellivue MP2 Firmware (CVSS 8.2, High).

No known exploitation. EPSS puts it in the 31st percentile. A vendor fix is available.

Check it yourself

No ready-made check. Neither Nuclei nor OpenVAS has one for this CVE yet.

Request a scan

Published Updated Sources: CVE.org, NVD, CSAF

What to do

The vendor has published a fix. Version details are below where the sources state them.

  • ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal MonitorsAffected: The following IntelliVue Patient Monitors versions are affected: IntelliVue…Upgrade to: not statedStated by: CVE record

Versions as the sources state them. Full version matrix after sign-in

What it is

From the CVE record

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.

In plain language

No plain-language summary for this CVE yet.

Exploitation

Where each signal puts this CVE on the scale from published to confirmed exploited.

EPSS31st
CISA KEV

Not in the catalog. CISA has not confirmed exploitation.

Public exploits

No public exploit or proof of concept found in the sources we track.

EPSS

0.4% chance of exploitation activity in the next 30 days, which ranks it in the 31st percentile of scored CVEs.

Exploit Prediction Scoring System, FIRST.org. A probability, not a confirmation.

Lifecycle

5 events over 2361 days, from the signal feeds we watch.

  1. Patch availablerecord updated
  2. Publishedweakness classified, att&ck mapped

Affected products

Technical detail

CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H

Scored 8.2 by NVD.

How it is reached

  • Attack Vector AdjacentRequires access to the local network (e.g. same Wi-Fi, Bluetooth)
  • Attack Complexity HighRequires specific conditions like a race condition or non-default configuration
  • Privileges Required NoneNo authentication required — anyone can exploit this
  • User Interaction NoneNo user interaction needed — fully automated exploitation

Scope

  • Scope ChangedThe exploit can affect other components (e.g. sandbox escape, host compromise from VM)

Impact if exploited

  • Confidentiality HighTotal information disclosure — all data in the component is compromised
  • Integrity LowData modification is possible but limited in scope or consequence
  • Availability HighTotal denial of service — the component is completely unavailable

Weaknesses

ATT&CK techniques

Mapped from the weaknesses above (CWE to ATT&CK), not observed in attacks.

Sources

Questions about CVE-2018-10601

Is CVE-2018-10601 being exploited?
No exploitation is confirmed and no public exploit is known. EPSS puts the chance of exploitation in the next 30 days at 0.4%.
Which versions are affected by CVE-2018-10601?
IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors: The following IntelliVue Patient Monitors versions are affected: IntelliVue….
How do I fix or mitigate CVE-2018-10601?
The vendor has published a fix. The advisory is on ics-cert.us-cert.gov.

Last updated

Watch the software you run.

My Stack ranks new CVEs for your products by real-world exploitation, so the next serious one reaches you without reading every advisory.

We'll flag the next CVE, public exploit or patch for Intellivue Mp2 Firmware, not every advisory.